Equipment designed for the administration of inhalation anaesthesia and/or patient ventilation during surgical or diagnostic procedures.

Description

Features

  • From small to large animals.
  • Electric piston ventilator (no motive gas consumption).
  • Very low system complexity due to piston ventilator operation.
  • Volume controlled ventilator mode (VCV) as standard and optional pressure-controlled mode (PCV), pressure support mode (PS), and synchronised intermittent ventilation (SIMV).
  • 2 gases, oxygen and air.
  • High precision dual-tube analogue rotameters.
  • Monitoring of ventilator volume and pressure curves on 6.5” display.
  • Tidal volume from 20 to 1400 ml (10 ml in pressure-controlled ventilation - PCV).
  • Inspiratory pressure from 5 to 65 cmH20.
  • Rate of 4 to 60 breaths per minute.
  • A patient system with a soda-lime canister allowing low flow rates.
  • Emergency oxygen.
  • Battery for at least 45 minutes.
  • Ventilator with the ability to continue ventilating without gas (electric) and/or without power supply, taking ambient air.
Practical implementation

All procedures requiring a ventilated animal (thoracic surgery).

Quality certifications

ISO 9001

Terms of use

To be used in an operating theatre equipped with oxygen gas and medical air intake under the supervision of a veterinary technician.