UV

TuMag Timelines Reference Sheet

SOPHI
Timestamp Action Commands Description Images Produced Expected outcome
00:00:00 Abort
00 HP Abort and FGReset
Abort & Reset 0
  • FG Reset (Instrument config.)
00:02:40 Initial Setting 01 Setting for ROI, Gain & Exp time 0
  • ROI (Instrument config.)
  • Gain (Instrument config.)
  • Exposure time (Instrument config.)
00:03:50:000 Prefilter scan 32 Scanning voltage range for the Mg+5250.2 filters 35/cam * 2 rep * 2
  • Thumbnails?
  • HVPS to -100V
  • FW to 517
  • HVPS from -100V to -3500V
  • HVPS from -3500 to -100V
  • HVPS to +100V
  • FW to 5250.2
  • HVPS from +100 to +3500V
  • HVPS from +3500V to +100V
  • HVPS to 0V
  • Images
00:04:50 Ping 99 Ping after CALL for sync 0
  • Pong
00:04:55 Ping 99 Ping after CALL for sync 0
  • Pong
00:06:00 Obs0p+2.02 13 Repetition of 5000 Obs 0p + 2.02 48/cam + 32/cam
  • Thumbnails.
  • Voltage changes from -3575V to -80V (Mode 0p) and +1730V to +2885V (Mode 2)
  • Images obatined
00:07:35 Stop Observation 97 0
  • Stop
00:08:05 Set voltage to 0V 96 0
  • Voltage to 0V
00:08:10 Ping 99 Ping after CALL for sync 0
  • Pong
00:08:15 Ping 99 Ping after CALL for sync 0
  • Pong