TuMag Timelines Reference Sheet

SOPHI
Timestamp Action Commands Description Images Produced Expected outcome
00:00:00 Abort
00 HP Abort and FGReset
Abort & Reset 0
  • FG Reset (Instrument config.)
00:02:40 Initial Setting 01 Setting for ROI, Gain & Exp time 0
  • ROI (Instrument config.)
  • Gain (Instrument config.)
  • Exposure time (Instrument config.)
00:04:50 Ping 99 Ping after CALL for sync 0
  • Pong
00:04:55_000 Prefilter scan 32 Scanning voltage range for the Mg+5250.2 filters 35/cam * 2 rep * 2
  • Thumbnails?
  • HVPS to -100V
  • FW to 517
  • HVPS from -100V to -3500V
  • HVPS from -3500 to -100V
  • HVPS to +100V
  • FW to 5250.2
  • HVPS from +100 to +3500V
  • HVPS from +3500V to +100V
  • HVPS to 0V
  • Images
00:07:05 Ping 99 Ping after CALL for sync 0
  • Pong
00:08:10_000 Obs Mod 0s + 3.02 17 Repetition of 5000 modes 0s+3.02 12/cam + 10/cam
  • Thumbnails arriving
  • Voltage changes from -3575V to -80V (Mode 0s) and +1860V to +2885V (Mode 3)
  • Images obatined
00:10:20_000 Stop Observation 97 Stops observation and empties queue 0
  • Stop current observation
00:12:30 Obs1+2.02 15 Repetition of 5000 Obs 1 + 2.02 40/cam + 32/cam
  • Thumbnails.
  • Voltage changes in the negative (Mode 1) and positive (Mode 2).
00:14:40 Ping 99 Ping after CALL for sync 0
  • Pong
00:16:50_000 Obs Mod 4 + 5.02 19 Repetition of 5000 modes 4+5.02 12/cam + 12/cam
  • Thumbnails arriving
  • Voltage changes from -2510V to -1980V (Mode 4) and +2000V to +2260V (Mode 5)
  • Images obatined
00:18:25_000 Stop Observation 97 Stops observation and empties queue 0
  • Stop current observation
00:18:55_000 Set HVPS to 0V 96 Sets HVPS to 0V 0
  • Voltage to 0V
00:19:00 Ping 99 Ping after CALL for sync 0
  • Pong
00:19:05 Ping 99 Ping after CALL for sync 0
  • Pong