UV

TuMag Timelines Reference Sheet

SOPHI

fs_1

Timestamp Action Commands Duration (s) Description Images Produced Expected outcome
00_00_00_00_000 Abort 00 120 Abort & reset 0
  • Stop all
  • Reset FG
00_00_01_50_000 Initial Settings 01 5 Settings for ROI, Gain & Exposure time 0
  • Changes in the cameras parameters
00_00_02_40_000 Stop Observation 97 30 Stops observation and empties queue 0
  • Stop current observation
00_00_03_10_000 Thumbnail Mg 51 1 Observation in Mg center for thumbnailing 1/cam
  • Thumbnails
  • Voltage to -2245V
  • Images
00_00_03_21_000 Abort 00 120 Abort & reset 0
  • Stop all
  • Reset FG
00_00_04_28_000 Initial Settings 01 5 Settings for ROI, Gain & Exposure time 0
  • Changes in the cameras parameters
00_00_04_38_000 Darks 31 139.3 Acquisition of darks 1/cam * 50
  • Thumbnails
  • HVPS at 0V
  • Images
00_00_09_05_000 Phase Diversity 33 28.8 PD measurements for 5250.2+Mg filters 40/cam * 2 (PD/F4) * 2
  • Thumbnails
  • HVPS to +2880V
  • FW to 5250.2 + PD
  • FW to 5250.2 + F4
  • HVPS to -80V
  • FW to 517 + PD
  • FW to 517 + F4
  • Images
00_00_11_42_000 Linear polarizer for 3 lines TUMAG with LCVR set 34 19.4 Linear polarizer measurements for the 3 prefilters 4/cam * 3
  • Thumbnails
  • FW to PL, 5250.2
  • HVPS to +2880V
  • FW to 517
  • HVPS to -80V
  • FW to 5250.6
  • HVPS to -1750V
  • HVPS to 0V
  • Images
00_00_12_12_000 Micropolarizers for 3 lines TUMAG with LCVR set 35 19.4 Micropolarizer measurements for the 3 prefilters 4/cam * 3
  • Thumbnails
  • FW to MP, 5250.2
  • HVPS to +2880V
  • FW to 517
  • HVPS to -80V
  • FW to 5250.6
  • HVPS to -1750V
  • HVPS to 0V
  • Images
00_00_12_42_000 Pinholes 36 11.4 Pinhole measurements in 3 filters 1/cam * 3
  • Thumbnails
  • FW to PH, 5250.6
  • HVPS to -1750V
  • FW to 517
  • HVPS to -80V
  • FW to 5250.2
  • HVPS to +2880V
  • HVPS to 0V
  • Images
00_00_13_12_000 FlatfieldsLong Obs0p 81 2202.5 Flat field observations for mode 0p (14 rep. obs, 4 lambda repeat) 48/cam * 4 lam_rep * 14 rep
  • Thumbnails
  • FW to 517
  • HVPS from -3580V to -80V
  • Images
  • HVPS to 0V
00_00_57_07_000 Stop Observation 97 30 Stops observation and empties queue 0
  • Stop current observation
00_00_58_32_000 FlatfieldsLong Obs2.02 82 1096.7 Flat field observations for mode 2.02 (11 rep. obs, 4 lambda repeat) 32/cam * 4 lam_rep * 11 rep
  • Thumbnails
  • FW to 5250.2
  • HVPS from +1729V to +2885V
  • Images
  • HVPS to 0V
00_01_21_22_000 Stop Observation 97 30 Stops observation and empties queue 0
  • Stop current observation
00_01_23_12_000 FlatfieldsLong Obs1 83 1813.3 Flat field observations for mode 1 (14 rep. obs, 4 lambda repeat) 40/cam * 4 lam_rep * 14 rep
  • Thumbnails
  • FW to 517
  • HVPS from -3243V to +85V
  • Images
  • HVPS to 0V
00_01_59_52_000 Stop Observation 97 30 Stops observation and empties queue 0
  • Stop current observation
00_02_15_37_000 Stop Observation 97 30 Stops observation and empties queue 0
  • Stop current observation
00_02_16_12_000 Thumbnail Mg 51 1 Observation in Mg center for thumbnailing 1/cam
  • Thumbnails
  • Voltage to -2245V
  • Images
00_02_16_17_000 Ping 99 0 Ping for timestamping 0
  • Pong
00_02_16_27_000 Ping 99 0 Ping for timestamping 0
  • Pong
00_02_16_28_000 Ping 99 0 Ping for timestamping 0
  • Pong
00_02_17_28_000 Obs Mod 2.02 21 29.5 Repetition of 5000 mode 2.02 32/cam
  • Thumbnails arriving
  • Voltage changes from +1730V to +2885V (Mode 2)
  • Images obatined
00_08_20_53_000 Stop Observation 97 30 Stops observation and empties queue 0
  • Stop current observation
00_08_21_23_000 Thumbnail Mg 51 1 Observation in Mg center for thumbnailing 1/cam
  • Thumbnails
  • Voltage to -2245V
  • Images
00_08_21_34_000 Abort 00 120 Abort & reset 0
  • Stop all
  • Reset FG
00_08_24_00_000 Initial Settings 01 5 Settings for ROI, Gain & Exposure time 0
  • Changes in the cameras parameters
00_08_24_22_000 Darks 31 139.3 Acquisition of darks 1/cam * 50
  • Thumbnails
  • HVPS at 0V
  • Images
00_08_29_27_000 Phase Diversity 33 28.8 PD measurements for 5250.2+Mg filters 40/cam * 2 (PD/F4) * 2
  • Thumbnails
  • HVPS to +2880V
  • FW to 5250.2 + PD
  • FW to 5250.2 + F4
  • HVPS to -80V
  • FW to 517 + PD
  • FW to 517 + F4
  • Images
00_08_32_08_000 Linear polarizer for 3 lines TUMAG with LCVR set 34 19.4 Linear polarizer measurements for the 3 prefilters 4/cam * 3
  • Thumbnails
  • FW to PL, 5250.2
  • HVPS to +2880V
  • FW to 517
  • HVPS to -80V
  • FW to 5250.6
  • HVPS to -1750V
  • HVPS to 0V
  • Images
00_08_56_28_000 FlatfieldsLong Obs2.02 82 1096.7 Flat field observations for mode 2.02 (11 rep. obs, 4 lambda repeat) 32/cam * 4 lam_rep * 11 rep
  • Thumbnails
  • FW to 5250.2
  • HVPS from +1729V to +2885V
  • Images
  • HVPS to 0V
00_09_32_00_000 Stop Observation 97 30 Stops observation and empties queue 0
  • Stop current observation
00_09_34_50_000 Micropolarizers for 3 lines TUMAG with LCVR set 35 19.4 Micropolarizer measurements for the 3 prefilters 4/cam * 3
  • Thumbnails
  • FW to MP, 5250.2
  • HVPS to +2880V
  • FW to 517
  • HVPS to -80V
  • FW to 5250.6
  • HVPS to -1750V
  • HVPS to 0V
  • Images
00_10_16_12_000 Pinholes 36 11.4 Pinhole measurements in 3 filters 1/cam * 3
  • Thumbnails
  • FW to PH, 5250.6
  • HVPS to -1750V
  • FW to 517
  • HVPS to -80V
  • FW to 5250.2
  • HVPS to +2880V
  • HVPS to 0V
  • Images
00_10_37_52_000 Prefilter scan 32 158.5 Scanning voltage range for the Mg+5250.2 filters 35/cam * 2 rep * 2
  • Thumbnails?
  • HVPS to -100V
  • FW to 517
  • HVPS from -100V to -3500V
  • HVPS from -3500 to -100V
  • HVPS to +100V
  • FW to 5250.2
  • HVPS from +100 to +3500V
  • HVPS from +3500V to +100V
  • HVPS to 0V
  • Images
00_10_47_32_000 Thumbnail Mg 51 1 Observation in Mg center for thumbnailing 1/cam
  • Thumbnails
  • Voltage to -2245V
  • Images
00_10_47_37_000 Ping 99 0 Ping for timestamping 0
  • Pong
00_10_47_47_000 Ping 99 0 Ping for timestamping 0
  • Pong
00_10_47_48_000 Ping 99 0 Ping for timestamping 0
  • Pong
Universitat de Valencia 2022