Monitor VetCare multiparamétrico Braun V9233B
Braun Multiparametric Monitor
Brand: Braun V9233B Multi-parametric VetCare Monitor
Type: Equipment

Equipment designed to monitor the patient's vital signs (ECG, pulse oximetry, capnography, invasive and non-invasive blood pressure, temperature).

Description

Display indicators and keys:

  • Main display:
    • 10.4-inch-high resolution active colour LCD.
    • Resolution 640 x 480 pixels.
  • Indicators:
    • Green LED and mains connection icon.
    • Green LED and battery operation icon.
  • Function keys:
    • On/Off, PI zero, start/stop PANI, start/stop the printer, alarm mute, rotary knob with push to select.

SpO2:

  • Range 0-100% (functional saturation).
  • Accuracy +/-2% between 70-100%, < 70% unspecified.
  • High and low alarms 50-100% and off.
  • Selectable average: slow (16 beats), normal (8 beats) or fast (4 beats).

Heart rate:

  • Range:
    • 20-350 bpm (SpO2).
    • 20-350 bpm (PANI) - Cat/Dog.
    • 40-180 bpm (PANI) - Horse.
    • 20-350 bpm (ECG).
  • Accuracy +/-2 bpm or +/-3% (whichever is greater).
  • Alarms 20-350 bpm and off.
  • Average Fixed at 8 seconds.

ECG:

  • 3 or 5 electrode configuration.
  • Electrodes I, II, III, aVR, aVL, aVF and V selection.
  • Gain X0.5, X1, X2, X4.
  • Defibrillator protection Yes.
  • Sweep speed 6.25, 12.5, 25 or 50 mm/s.

Non Invasive Blood Pressure (NIBP):

  • Range 20-250 mmHg (Cat/Dog/Horse).
  • Accuracy is Equal to or better than specified in the SP10-1992 AAMI standard for PANI accuracy by oscillometric method.
  • Inflation pressure 50-250 mmHg (Cat) 50-200 mmHg (Dog/Horse) Default inflation pressure 200 mmHg (Cat), 150 mmHg (Dog), 100 mmHg (Horse).
  • Alarm range 20-250 mmHg and off.

Invasive Pressure 2 channels:

  • Range -50 to 300 mmHg (after zeroing).
  • Accuracy +/-1 mmHg or +/-1% (whichever is greater).
  • Alarms -50 to 300 mmHg.

Temperature 2 channels:

  • Range 0-50º C.
  • Accuracy +/-0,1º C plus temperature sensor tolerance.
  • Resolution 0,1º C.
  • Alarms 20.0-50.0º C and off.
  • YSI 400 Series sensor compatibility.

Sidestream Capnography:

  • Measurement method non-dispersive IR absorption.
  • Measurement range 0-10% CO2 STPD.
  • Display range 0-100 mmHg; 0-13.3 kPa; 0-10% CO2.
  • Accuracy +/-2 mmHg or 4% of reading (whichever is greater) (0-10% CO2).
  • Alarms 0-100 mmHg; 0-13.3 kPa; 0-10% CO2 and off.
  • N2O compensation Selectable 40% (off by default).
  • Average 4 breaths.
  • Flow 150 ml/min +/-20 m.
Practical implementation
  • Patient vital signs monitoring.
Quality certifications

ISO 9001

Terms of use

Only suitable for animals whose determinations are within the range of use (not suitable for rodents).