Type: Equipment
High-performance Quadrupole Time-of-Flight (QTOF) mass spectrometer for accurate mass and high-resolution analysis. This system is essential for identifying unknown compounds and performing advanced quantification studies in complex samples.
-
Sensitivity and Resolution:
-
TOF/MS Sensitivity: in positive ESI mode for a 1 pg on-column reserpine injection, producing a S/N ratio > 2500.
-
MS/MS Sensitivity: in positive ESI mode for a 50 fg on-column reserpine injection, producing a S/N ratio > 400.
-
TOF/MS Sensitivity with Twin Sprayer Probe: 1 pg on-column reserpine injection produces S/N > 750.
-
MS Resolution: higher than 42,000 (FWHM) m/z 956.
-
-
Performance and Acquisition:
-
MS/MS Scan Speed: up to 133 Hz (133 MS/MS spectra per second).
-
MS/MS duty cycle: ≥ 90% for TOF MS/MS acquisition.
-
Scanning modes: Full scan TOF MS, Full scan TOF MS/MS, MRMHR, SWATH (DIA), IDA (DDA).
-
Linear dynamic range: better than 5 orders of magnitude without signal loss.
-
-
Accuracy and Ranges:
-
Internal mass accuracy: less than 0.2 ppm absolute error.
-
External mass accuracy: less than 2 ppm RMS over 12 hours of LC-MS operation.
-
Mass range: up to 40 kDa in TOF; Precursor Ion 5-2,250 m/z (5-40,000 m/z).
-
Workflow flow range: from 5 μL/min to 3 mL/min.
-
Fragmentation modes: CID collision cell and EAD cell.
-
-
Identification of molecular formulas and structural characterization of complex molecules.
-
Metabolomics and proteomics studies using advanced acquisition modes (SWATH, IDA).
-
Analysis of large molecules and biotherapeutics up to 40 kDa.
Certificat Buerau Veritas ISO9001
10-1-2020 al 2023
The use of the equipment will only be carried out by the technical staff of the section and authorized users for the operation of the equipment.
- Galletero Pedroche, M.sales
- PTGAS-Ets Investigacio-Espectrom. Mases, Rmn
- Lliso Escudero, Isabel
- PIT-Tecnic/a Sup Uv
- Martinez Agudo, Jose Maria
- PTGAS-Etm Investigacio-Biolog.,Quim.,o Sanit.