Mass spectrometer QTOF 6600+ and UHPLC
Brand: TRIPLETOFT6600+/Exion LC AD Pump, by SCIEX
Type: Equipment
Type: Equipment
High resolution hybrid mass spectrometer and quadrupole time-of-flight analyser TRIPLETOF 6600+ and UHPLC (QTOF6600+).
Description
QTOF 6600
- High sensitivity (MS and MS/MS)
- Acquisition speed of 100 Hz; High acquisition speed without compromising resolution of data to identify more compounds in the same sample
- High resolution (≥35000 (FWHM) at 956 m/z
- Mass accuracy (~ 1 ppm RMS)
- Linear dynamic range (~4 orders RMS)
- Hybrid triple quadrupole time-of-flight system. Specifications typical of fast, high resolution and sensitive equipment, without compromising any of the characteristics. High versatility in identification and quantification.
- The system is composed by the following parts:
- ionisation source: ESI/APCI
- Interface between the ionisation source and analysers
- quadrupole analysers
- packing system and exclusive ion acceleration
- TOF: High capacity, resolution and speed time-of-flight analyser
- Detector: Wide range dynamic multi-canal TDC
- Liquid chromatography system Exion LC AD
- Constant flow and pressure (10-1300 bar)
- High pressure gradient
- Pressure measurement accuracy. Better than ±2% or 10 bar
- Built-in sample refrigeration system between 4ºC and 40ºC.
- Volume injection precision. RSD ≤ 0.25% ;
- Volume injection precision. ±1%
- Cross-contamination. Maximum of 0.0015%
- Vacuum speed. From 0.1 to 5 μL /s (in increments of 0.1 μL /s)
- Highest performance pressure, up to 1300 bar
- Thermostat column oven from 4ºC to 150ºC
UHPLC
Key features of the mass spectrometer:
- Exion LC™ AD High pressure pump:
- Constant flow and pressure and high pressure gradient
- Dual piston micro-volume pumps
- High pressure gradient system
- Exion LC degasser
- High pressure automatic sampler
- Direct injection method of variable volume. Without sample loss during injection.
- Built-in sample refrigeration system between 4ºC and 40ºC.
- Injection volume. From 0.1 to 50 μL. 0.1 to 0.9 μL in increments of 0.1 μL, 1 a
- Sample capacity: 96 y 384 μL vials and plates
- Volume injection precision. RSD ≤ 0.25%
- Volume injection precision. ±1%
- Column oven
- Temperature control from 4ºC to 150º above room temperature
- Solvent tray
Practical implementation
QTOF 6600
The equipment includes the following programmes:
- PeakView: quick and efficient identification of compounds (own databases: pesticides, antibiotics-anthelmintics, organophosphates, nitrosamines, perfluorates, steroids...)
- MultiQuant: to carry out procedures at triple quadrupole level.
- Markerview: metabonomics for the analysis of main compounds and statistical study of results of compound and metabolite identification
- MetabolitePilot: metabolite identification
- LipidView: Lipids analysis
UHPLC
- Compound separation for high resolution mass spectrometer detection (QTOF application).
Quality certifications
Bureau Veritas ISO9001 Certificate
From 10-1-2020 to 2023
Terms of use
Only the technical staff from the handling section will use the equipment.
UV structures that manages it
Central Service for Experimental Research (SCSIE)
Mass spectrometry